MELAG Premium, B-Class
![]() The chamber offers sufficient place for the goods to be sterilized |
![]() Vacuklav 40-B with assembled MELAdem 40 and MELAjet |
![]() For the surgery: Sterilzation in sterilization containers |
![]() Mount >C< for 6 Trays (standard) |
![]() Mount >C< (after being turned) for 3 Standard-tray-cassettes |
![]() Mount >B< for 4 Standard-tray-cassettes (or for 4 trays) |
![]() Mount >D< for 2 higher cassettes (max. Dimensions: 19 x 29 x 7 cm, i.e. implantology trays) |
![]() Installation scheme for Vacuklav 40-B with MELAdem 40 and MELAjet |
![]() Built-in CF-Slot for the documentation with the MELAflash CF-Card |
![]() |
|
||||||||
The Vacuklav works with a fractionated pre-vacuum and satisfies the requirements of “Class B”- autoclaves in the European standard EN 13060. The deep vacuum is created by a very high performance electrical vacuum pump. The deep post-vacuum ensures optimal drying of wrapped materials, large quantities of wrapped textiles and porous goods. The Vacuklav 40-B and 43-B can be connected to a water treatment system, to a water source and a drain. The tap water will be used to cool the vacuum pump. The Vacuklav 41-B and 44-B can be connected with a water treatment system and drain as an option. SPEED COMFORT DESIGN
|
|||||||||
| Technical Data | ||
|---|---|---|
| Dimensions | Vacuklav 40-B, 41-B | Vacuklav 43-B, 44-B |
| External dimensions: | Width: 46 cm, Height (incl. display): 56 cm, Depth: 57 cm | Width: 46 cm, Height (incl. display): 56 cm, Depth: 68 cm |
| Chamber: | Diameter 25 x 35 cm | ø 25 x 45 cm |
| Volume of the chamber: | 18 litres | 22 litres |
| Tray: | Width: 19 cm, Depth: 29 cm, Height: 2 cm, | Width: 19 cm, Depth: 42 cm, Height: 2 cm, |
| Weight: | 56 kg (40-B) 58 kg (41-B) |
57 kg (43-B) 59 kg (44-B) |
| Power Supply: | 230 Volt / 50 Hz / 3,400 Watt | 230 Volt / 50 Hz / 3,400 Watt |
| Program Cycle Times | |||
|---|---|---|---|
| Program | Cycle time (warm start): | Sterilization time: | Drying time: |
| Universal-Program wrapped instruments (134°C / 2 bar) |
approx. 15 minutes | 5 minutes | 15 minutes |
| Quick-Program„B“ single wrapped instruments (max. 1,5 kg) (134°C / 2 bar) |
approx. 11 minutes | 3.5 minutes | 6 minutes |
| Quick-Program „S“ unwrapped instruments (134°C / 2 bar) |
approx. 8 minutes | 3.5 minutes | 1 minutes |
| Gentle-Program wrapped instruments (121°C / 1 bar) |
approx. 27 minutes | 20 minutes | 15 minutes |
| Prion-Program wrapped instruments (134°C / 2 bar) |
approx. 30 minutes | 20 minutes | 15 minutes |
| Bowie&Dick-test (134°C / 2 bar) |
approx. 15 minutes | 3.5 minutes | 15 minutes |
| Vacuum-test | approx. 16 minutes | - | - |
![]() Quick-Program (134°C / 2 bar) |
![]() Universal-Program (134°C / 2 bar) |
![]() Gentle-Program (121°C / 1 bar) |
![]() Prion-Program (134°C / 2 bar) |